SLAOT 2016

SLAOT 2016
Punta Cana, República Dominicana

lunes, 14 de octubre de 2013

Metal ion levels from well-functioning Birmingham Hip Resurfacings decline significantly at ten years

http://www.bjj.boneandjoint.org.uk/content/95-B/10/1332.abstract

Metal ion levels from well-functioning Birmingham Hip Resurfacings decline significantly at ten years

  1. K. De Smet, MD, Orthopaedic Surgeon2
+Author Affiliations
  1. 1Ghent University Hospital, Department of Orthopaedics and Traumatology, De Pintelaan 185, 9000 Ghent, Belgium.
  2. 2ANCA Medical Centre, Xavier De Cocklaan 68/1, B-9831 St-Martens-Latem, Belgium.
  3. 3Anca Clinic Roma, Via G. De Notaris 2/B, 00197 Roma, Italy
  1. Correspondence should be sent to Dr C. Van Der Straeten; e-mail:Catherine.VanDerStraeten@UGent.be

Abstract

A retrospective study was conducted to investigate the changes in metal ion levels in a consecutive series of Birmingham Hip Resurfacings (BHRs) at a minimum ten-year follow-up. We reviewed 250 BHRs implanted in 232 patients between 1998 and 2001. Implant survival, clinical outcome (Harris hip score), radiographs and serum chromium (Cr) and cobalt (Co) ion levels were assessed.
Of 232 patients, 18 were dead (five bilateral BHRs), 15 lost to follow-up and ten had been revised. The remaining 202 BHRs in 190 patients (136 men and 54 women; mean age at surgery 50.5 years (17 to 76)) were evaluated at a minimum follow-up of ten years (mean 10.8 years (10 to 13.6)). The overall implant survival at 13.2 years was 92.4% (95% confidence interval 90.8 to 94.0). The mean Harris hip score was 97.7 (median 100; 65 to 100). Median and mean ion levels were low for unilateral resurfacings (Cr: median 1.3 µg/l, mean 1.95 µg/l (< 0.5 to 16.2); Co: median 1.0 µg/l, mean 1.62 µg/l (< 0.5 to 17.3)) and bilateral resurfacings (Cr: median 3.2 µg/l, mean 3.46 µg/l (< 0.5 to 10.0); Co: median 2.3 µg/l, mean 2.66 µg/l (< 0.5 to 9.5)). In 80 unilateral BHRs with sequential ion measurements, Cr and Co levels were found to decrease significantly (p < 0.001) from the initial assessment at a median of six years (4 to 8) to the last assessment at a median of 11 years (9 to 13), with a mean reduction of 1.24 µg/l for Cr and 0.88 µg/l for Co. Three female patients had a > 2.5 µg/l increase of Co ions, associated with head sizes ≤ 50 mm, clinical symptoms and osteolysis. Overall, there was no significant difference in change of ion levels between genders (Cr, p = 0.845; Co, p = 0.310) or component sizes (Cr, p = 0.505; Co, p = 0.370). Higher acetabular component inclination angles correlated with greater change in ion levels (Cr, p = 0.013; Co, p = 0.002). Patients with increased ion levels had lower Harris hip scores (p = 0.038).
In conclusion, in well-functioning BHRs the metal ion levels decreased significantly at ten years. An increase > 2.5 µg/l was associated with poor function.
Cite this article: Bone Joint J 2013;95-B:1332–8.

No hay comentarios:

Publicar un comentario